Operation
SV 125
4016589
5--104
Version 1.0
Aeration:
Duration of Aeration [ms]
The venting time is entered
(To prevent syringes from sticking to the vacuum chamber after vacuum filling,
venting with nitrogen is performed after the vacuum filling).
Aeration stop [ms]
Aeration ends after the entered time.
Delay aeration on [ms]
Delay with which aeration is switched on.
Delay aeration off [ms]
Delay with which aeration is switched off.
Leaktest limit difference [mbar]
The limit value that the pressure is allowed to drop during leak test is entered.
Min. vacuum tank filling [mbar]
Entry of the required minimum pressure that must be reached in the vacuum tank,
before the vacuum valve opens.
Actual vacuum tank filling [mbar]
Display of the current pressure in the vacuum tank.