INNOWATER PHRH
The value on the left (
Lec:)
: indicates the current ORP (RH) value measured by the
probe. The value on the right, indicates the value of the buffer solution (220mV).
You can adjust this value using the arrows to adapt it to the temperature and the
sample used.
Insert the probe in the buffer solution 220mV, remove it slightly with the probe and
wait for a stable reading value to be reached.
Once the reading value has stabilized press
OK
key to save the calibration and exit
the sub
-
menu.
Next, remove the probe from the buffer solution 220mV, rinse its bottom with clean
water and shake gently to remove excess water.
5.4 Factory calibration
The submenu
Factory calib
, gives you the possibility to reset general calibration
parameters that correspond, approximately, with those of a new probe and which
are the ones programmed by the chlorinator from factory values. This can be useful
if you have saved successive calibrations and do not have the buffer solutions for a
correct calibration.
Press
OK
key to enter the sub
-
menu
5 Factory calib
, it will show the following
screen:
Press
OK
key to set Factory values or
MENU
to Exit.
6. Language
Press
OK
key to enter and scroll with the help of the arrows to choose the desired
language.
Press
OK
key to confirm and Exit.
7. Contrast LCD
Press
OK
key to enter and scroll with the help of the arrows to choose the desired
LCD contrast.
Press
OK
key to confirm and Exit.
MAIN MENU
3 Language
Choose language
English
OK
ORP Config.
3 Factory Cal.
Factory Calib.?
SI:OK EXIT:MENU
OK
LCD Contrast
-
■ ■ ■ ■ +
MAIN MENU
4 LCD Contrast