15
Activate external control:
In
“2
mode
”,
choose the option
“
Open = active
”:
In
“3
Production
”,
set 0%:
This way, when chlorinator receives the order to produce chlorine, it will do it at the
percentage chosen from the main screen. If the order is not to produce, it will
remain at 0%.
5.3 Probe calibration
The ORP probes require a calibration before their first use and then they need to be
calibrated periodically. It is because different probes can have different answers and
because the response of the same probe inevitably varies with time.
The calibration consists of measuring the response of the probe introducing it in two
buffer solution and recording this response in order to deduct the ORP (RH) of any
other solution, in our case, the ORP (RH) of the water in the pool.
The calibration of the probe is carried out using a buffer solution supplied (220mV)
and entering the sub
-
menu
"3 ORP Calib.
”
.
Enter the sub
-
menu
3 ORP Calib.
by pressing
OK
key, it will show the following
screen:
MENU CONTROL EXT
1 on /off
External control
ACTIVATED
OK
MENU CONTROL EXT
2 Mode
Mode
Open = active
OK
MENU CONTROL EXT
3 Production
Production if
active 0%
OK
ORP configuration
3 Cal ORP
ORP Calibration
Lec: 400mV 220