19P03-DOC01E REV-3
85
CPR handle replacement
Tools required:
•
#2 Robertson Screwdriver
•
9/16” Combination Wrench
Procedure:
1. Raise the bed to its highest
position and apply the
brakes.
2. Extend the tight section to its
max length
.
3. Unplug the power cord of the
bed from the wall outlet.
4. With a 9/16’’ combination wrench, unscrew the counter-nut (A) and then unscrew
completely the screw (B).
5. Replace the broken handle by the new one.
6. Reverse the above steps to fix the new handle in place
7. Verify that the CPR is working properly before returning the bed to service.
A
B