19P03-DOC01E REV-3
35
Alarm options
The system has two (2) types of basic alarms:
audible indication
and
visual indication
.
The audible warning will make a sound when the system detects a bed exit or bed exit
attempts (if option 2 zone is chosen). You may activate or deactivate this audible warning
according to your needs. To do so, press and hold the activating button of the zone 1 of
the bed exit alarm system until you hear one of the following audible signals:
1 beep: audible warning
deactivated
3 beeps: audible warning
activated
The visual indication will be shown by an indicator light located on each side of the foot
of the bed. A
red
indicator shows that the system is activated. This indicator will flash if
the system detects a bed exit or a bed exit attempt (if option 2 zone is chosen).
See image below:
Connection options to the nurse call system
The system is equipped with a junction box that may be connected to the nurse call
system. To connect it, your nurse call system must be able to interface with a DB37
connexion.
This is a standard type of connexion in the industry. The receptacle is located under the
top section of the bed structure at the end of the head section, right behind the trapeze bar
receptacle. If another type of junction is required, please contact us for available options.