19P03-DOC01E REV-3
83
Nurse Call replacement
Tools required:
•
Phillips Screwdriver
•
Trim clip removal tool
• 5/16
” Combination Wrench
•
Cutting Plier
•
3/8
” Combination Wrench
Procedure:
1. Raise the bed to its highest position and apply the brakes.
2.
Raise the backrest to its highest position and remove the headboard
.
3. Unplug the power cord of the bed from the wall outlet.
4. Using the rim clip removal tool, remove the plastic rivets and the protection plate.
5. Using the cutting pliers, remove the tie-wraps securing
the nurse call cable to the frame.
6. Press on the circuit lock and move the circuit left to
remove it.
CIRCUIT LOCK