Innomed Medical
Heart Screen 112C-1
Service manual
5/47
D-7443-1-C
Lead groups
Standard:
3 channels: I-III; aVR-aVF; V1-V3; V4-V6
6 channels: I – aVF; V1 – V6
12 channels: I…V6
Cabrera:
3 channels: aVL I -aVR; II aVF III; V1- V3; V4-V6
6 channels: aVL I –aVR II aVF III; V1 V2 V3 V4 V5 V6
12 channels: aV…V6
Heart frequency measurement
Measurement range
30 – 300 BPM
Accuracy
1 BPM or
2%
ECG amplifier
Input signal range
10 mV
Frequency response
0,05 Hz…150 Hz +0,-3 dB
Voltage measurement relative error
in the 0,5 - 10 mV range
3%
Sensitivity
2.5, 5, 10, 20 mm/mV Auto
Noise referred to the input
max. 20
Vpp
Input impedance (informative)
min. 20 MOhm
CMRR
min. 100 dB
DC tolerance
min.
320 mV DC
Pacemaker tolerance
max. 700mV/2ms
Pacemaker detection
min.2 mV/0,5 ms max. 700mV/2ms
Time constant
min. 3,2 sec
Patient leakage current
max. 10
A
Electrode failure detection
by channels
Calibrating signal (software generated)
1 mV / 300 ms
Muscle filter (digital)
35 Hz +0/-3 dB
Line filter (digital)
50 / 60 Hz
Line filter - 20 dB point:
50 Hz filter: 48,5 … 51,5 Hz
60 Hz filter: 59 … 61 Hz
Baseline filter (digital)
under 0,3 Hz: complete suppression
over 0,9 Hz: complete transfer
Amplifier error messages
Audio and visual signal in case of
input level error and noisy signal
Patient cable type
Innomed approved patient cable
Defibrillator protected 10-lead cable
4-lead cable
Defibrillator tolerance
Only with defibrillator protected cable!
max. 360 J, 5000 V
Digitalizing features
Data acquisition
12-lead, simultaneous (within 100us)
Sampling frequency
8000 Hz / channel
A/D resolution
24 bit, 0.076uV/bit
Nominal processed signal range
700 mV
Stored signal resolution
500 Hz 10 uV/bit
Resolution in case of PC connection during online ECG
transfer
500 Hz 5
V/bit
Printing features
Effective recording width per channel
40 mm, overlap allowed
Printer head resolution
In amplitude direction: 8 points / mm
On the time axis: 8 or 16 points / mm
Summary of Contents for Hear Screen 112C-1
Page 31: ...Innomed Medical Heart Screen 112C 1 Service manual 31 47 D 7443 1 C BOTTOM...
Page 33: ...Innomed Medical Heart Screen 112C 1 Service manual 33 47 D 7443 1 C BOTTOM...
Page 35: ...Innomed Medical Heart Screen 112C 1 Service manual 35 47 D 7443 1 C...
Page 36: ...Innomed Medical Heart Screen 112C 1 Service manual 36 47 D 7443 1 C...
Page 37: ...Innomed Medical Heart Screen 112C 1 Service manual 37 47 D 7443 1 C...
Page 38: ...Innomed Medical Heart Screen 112C 1 Service manual 38 47 D 7443 1 C...
Page 39: ...Innomed Medical Heart Screen 112C 1 Service manual 39 47 D 7443 1 C...
Page 40: ...Innomed Medical Heart Screen 112C 1 Service manual 40 47 D 7443 1 C...
Page 41: ...Innomed Medical Heart Screen 112C 1 Service manual 41 47 D 7443 1 C...
Page 42: ...Innomed Medical Heart Screen 112C 1 Service manual 42 47 D 7443 1 C...
Page 43: ...Innomed Medical Heart Screen 112C 1 Service manual 43 47 D 7443 1 C...
Page 45: ...Innomed Medical Heart Screen 112C 1 Service manual 45 47 D 7443 1 C...
Page 46: ...Innomed Medical Heart Screen 112C 1 Service manual 46 47 D 7443 1 C...