Innomed Medical
Heart Screen 112C-1
Service manual
19/47
D-7443-1-C
6.2.3 Latch-up watch check
Only connect the DC generator to the amplifier input according to the table. The DC level has to be controlled
between 0 and 450 mV. The filter must be turned off.
Connect the patient cable according to the following table and start the channels one-by-one with the channel
selector. On the display check the measurement order quality, the baseline, which can be reset after DC
controlling with the blocking command.
During the examination the device indicates the latch-up watch circuit’s operation with text message. During the
measurement enter the DC values into the table where the text appeared. The comparison levels must be min.
340 mV, max 450mV.
P1
P2
Examined lead
Positive DC (mV)
Negative DC (mV)
L
all other
I
R
all other
II
V1
all other
V1
V2
all other
V2
V3
all other
V3
V4
all other
V4
V5
all other
V5
V6
all other
V6
6.2.4 Measuring noise on the input and baseline checking
Make a 10 seconds long recording with all inputs connected together, at *2 sensitivity, 25 mm/sec speed from all
groups. The filter must be off. Measure the noise and absolute deviation from the baseline on the null-line.
The noise on the input is acceptable if the value of the noise peak-to-peak is not greater than
(0.4 mm+1 pixel)=0.525 mm on any lead. The null-line is acceptable if the deviation from the baseline is less
than 2 mm anywhere on the printout and on the display.
Summary of Contents for Hear Screen 112C-1
Page 31: ...Innomed Medical Heart Screen 112C 1 Service manual 31 47 D 7443 1 C BOTTOM...
Page 33: ...Innomed Medical Heart Screen 112C 1 Service manual 33 47 D 7443 1 C BOTTOM...
Page 35: ...Innomed Medical Heart Screen 112C 1 Service manual 35 47 D 7443 1 C...
Page 36: ...Innomed Medical Heart Screen 112C 1 Service manual 36 47 D 7443 1 C...
Page 37: ...Innomed Medical Heart Screen 112C 1 Service manual 37 47 D 7443 1 C...
Page 38: ...Innomed Medical Heart Screen 112C 1 Service manual 38 47 D 7443 1 C...
Page 39: ...Innomed Medical Heart Screen 112C 1 Service manual 39 47 D 7443 1 C...
Page 40: ...Innomed Medical Heart Screen 112C 1 Service manual 40 47 D 7443 1 C...
Page 41: ...Innomed Medical Heart Screen 112C 1 Service manual 41 47 D 7443 1 C...
Page 42: ...Innomed Medical Heart Screen 112C 1 Service manual 42 47 D 7443 1 C...
Page 43: ...Innomed Medical Heart Screen 112C 1 Service manual 43 47 D 7443 1 C...
Page 45: ...Innomed Medical Heart Screen 112C 1 Service manual 45 47 D 7443 1 C...
Page 46: ...Innomed Medical Heart Screen 112C 1 Service manual 46 47 D 7443 1 C...