Innomed Medical
Heart Screen 112C-1
Service manual
16/47
D-7443-1-C
5 Board level measurements
5.1 MAIN board volage test:
5.1.1 Checking voltages
– digital circuits
Check d5V voltage at test point J3/1 5V +/-5%
Check Vref +2.5V voltage at test point J3/7 5V +/-2%
5.1.2 Checking voltages
– Analog part
Check VCCAN buffer voltage at capacitor C109 6V +/-5%
Check analogue power supply 3.3V at JP5
3.3V +/-5%
Check Ref_Master voltage 1.8V at C45
1.8V +/-2%
Check Ref_Slave voltage 1.8V at C74
1.8V +/-2%
5.2
Voltage tests on the “interface” board
5.2.1 Checking voltages
Check voltage on J3/6 pontján:
+3,3V +/-5%
Check HEAD VOLTAGE during printing on the two
poles of C23
+7,8V +/-5%
Summary of Contents for Hear Screen 112C-1
Page 31: ...Innomed Medical Heart Screen 112C 1 Service manual 31 47 D 7443 1 C BOTTOM...
Page 33: ...Innomed Medical Heart Screen 112C 1 Service manual 33 47 D 7443 1 C BOTTOM...
Page 35: ...Innomed Medical Heart Screen 112C 1 Service manual 35 47 D 7443 1 C...
Page 36: ...Innomed Medical Heart Screen 112C 1 Service manual 36 47 D 7443 1 C...
Page 37: ...Innomed Medical Heart Screen 112C 1 Service manual 37 47 D 7443 1 C...
Page 38: ...Innomed Medical Heart Screen 112C 1 Service manual 38 47 D 7443 1 C...
Page 39: ...Innomed Medical Heart Screen 112C 1 Service manual 39 47 D 7443 1 C...
Page 40: ...Innomed Medical Heart Screen 112C 1 Service manual 40 47 D 7443 1 C...
Page 41: ...Innomed Medical Heart Screen 112C 1 Service manual 41 47 D 7443 1 C...
Page 42: ...Innomed Medical Heart Screen 112C 1 Service manual 42 47 D 7443 1 C...
Page 43: ...Innomed Medical Heart Screen 112C 1 Service manual 43 47 D 7443 1 C...
Page 45: ...Innomed Medical Heart Screen 112C 1 Service manual 45 47 D 7443 1 C...
Page 46: ...Innomed Medical Heart Screen 112C 1 Service manual 46 47 D 7443 1 C...