Declaration-Electromagnetic Immunity(Home Healthcare Environment)
Immunity test
IEC 60601 test level
Compliance level
Conducted RF
IEC 61000-4-6
3V
150 kHz to 80 MHz
6V in ISM and amateur
radio bands between
0.15MHz and 80MHz
3V
150 kHz to 80 MHz
6V in ISM and amateur
radio bands between
0.15MHz and 80MHz
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
also meet the requirement
of table 9 of 60601-1-2:2014
10 V/m
80 MHz to 2.7 GHz
also meet the requirement
of table 9 of 60601-1-2:2014
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV,±4 kV,±8 kV,±15 kV air
±8 kV contact
±2 kV,±4 kV,±8 kV,±15 kV air
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±2 kV for power supply
lines
Surge
IEC 61000-4-5
±0.5 kV,± 1 kV line(s) to lines
±0.5 kV,± 1 kV line(s) to lines
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
0% UT, 0.5 Cycle at 0°,45°,90°,
135°,180°, 225°, 270°, 315°
0% UT, 1 Cycle and 70% UT,
25/30 cycles
sigle phase:at 0°
0% UT,250/300 cycles
0% UT, 0.5 Cycle at 00,45°,90°,
135°,180°, 225°, 270°, 315°
0% UT, 1 Cycle and 70% UT,
25 cycles
sigle phase:at 0°
0% UT,250cycles
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m