7. Since neither the device nor batteries are household waste, follow your local
re cycling rules and dispose them at an appropriate collection site.
8. Do not open the device, or delicate electrical components a san intricate
air unit could be damaged. If you can not fix the problem using the
TROUBLESOOTING INSTRUCTION, please request service from your dealer.
WARNING: Do not repair the device without manufacturer’s authorization. Do
no carry out maintenance when using the device.
Caution: Generally, we recommend having the device inspected every 2 years
to ensure proper function, accuracy and safety. Please contact your dealer for
maintenance.
TROUBLESHOOTING
SYMPTOM
CHECK POINT
REMEDY
No display after pressing the
button
The batteries have run down.
The polarity of installed
batteries is wrong.
The contact of battery
compartment is polluted.
Replace all the batteries with
new ones.
Install the batteries correctly.
Clean the battery terminals
with dry cloth.
Inflation stops and re-inflate
later.
190mmHg is not enough
or the device has moved
excessively.
Did you talk or move your
arm (or hand) during
measurement?
See AUTOMATIC INFLATION.
Keep quit and silent during
the measurement.