CAN Infrascanner Model 2000 Operator Manual
INFRASCAN, INC. VERSION 1 Page 34 of 41
5.9 Error Messages.
Must be cleared by pressing the GREEN navigation button.
Error Message
Causes
High Signal.
Caused by too much ambient light, Infrascanner
is especially sensitive to infrared lamps or bright
incandescent bulbs. Try shielding the light
guide area with hands and repeat the
measurement.
Low Signal.
Hair may be trapped under the light guide or light
guide might not be making contact with scalp,
Repeat the measurement.
Repeat Measurement.
The light on one side of the patient’s head is
brighter than the other and the amplifiers need to
begin the measurement with a lower gain. Re-
take the data of the pair.
Saturated, Start in Reverse
Order.
The Infrascanner has detected a potential
hematoma and needs to re-take the data in the
reverse order (right side first) to confirm. The
message always occurs when a hematoma is
present on the left side (first side of pair to be
measured).
Timeout.
Infrascanner gains unable to converge quickly
enough. Move the light guides 5mm and repeat
the measurement.
Unstable Signal.
Caused by variations in the signal level. Repeat
the measurement at the site taking care to hold
Infrascanner steady.
Communications Error.
Software error. Take off the Disposable Fiber
Optic Shield and put it back on to reset the
Infrascanner.
Battery is Low.
Charge the Infrascanner
or replace the AA
disposable batteries.