24/ 25
Performing Your Test
Call toll free :
1-888-446-3246
www.infopiausa.com
Unscrew the lancing device cap.
Insert a sterile lancet into the lancing
device.
Twist the protective cover off.
Do not discard.
Replace the lancing cap by twisting it
back on tightly.
To adjust the depth setting:
Use 1-2 for soft skin, 3 for average skin,
4-5 for thick or calloused skin.
Performing Your Test
A blinking test strip will appear at the top of the
screen indicating the meter is ready for testing.
“Please apply blood to the tip of the test strip.”
Caution:
• Inserting the test strip in the wrong direction will turn the meter on but testing
will not be possible.
• If the meter does not power on, pull the test strip out of the port and reinsert
the test strip.
• Be careful not to bend the test strip when inserting it into the port.
• For a more accurate test result avoid testing under direct sunlight.
Warning:
• Element™ Plus meter and reusable lancing device are for single patient use
only.
Do not share them with anyone including other family members!
Do not use on multiple patients!
• All parts of the kit are considered bioharzardous and can potentially transmit
infectious disease, even after you have performed cleaning and disinfection.
• Avoiding getting hand lotion, oil, dirt or debris in or on the lancets and the reusable
lancing device.
• Please refer to page 40~45 for cleaning and disinfection Element™ Plus system.
• Wash hands thoroughly with soap and water after handling the meter, lancing device
or test strips.
Step 2
To obtain a suitable blood sample:
Before obtaining a blood sample wash your hands with warm water and
soap. Dry your hands completely before testing. If needed massage
your finger tip.