4 - 15
XTC/3 Operating Manual
to a well controlled rate of evaporant flux. If the required rate
control accuracy cannot be achieved within sixty seconds,
the
Delay Failure
alarm message will appear. If
Stop on
Alarms
has been set to
Yes
, the layer will
STOP
at the
same time.
Control. . . . . . . . Control Delay
state. The
Deposit
state is entered and the
Control Delay
suspends control loop action on the source
control power for the time interval programmed in
Control
Delay Time.
Source and sensor shutters are activated during
Control
Delay
. After the
Control Delay Time
elapses, control loop
action is initiated.
Both
. . . . . . . . . . XTC/3 will first enter a
Shutter Delay
state followed by
Control Delay
.
TRANSFER SENSOR
. . . . . . . . . . . Yes / No
Press the
TOGL
key to move through the choices. Default is
No
. This feature can
only be enabled when the
Delay
option
Shutter
or
Both
has been enabled. In
addition, a second sensor must be available to use this feature. If set to
Yes
, XTC/3
will use the second sensor for rate control during the shutter delay time and revert
back to the primary sensor after the shutter delay time has expired.
TRANSFER TOOLING
. . . . . . . . . . . 10.0 to 500.0%
This parameter is enabled if
Transfer Sensor
has been set to
Yes
. Sets the tooling
factor to use with the
Transfer
sensor. See
Tooling
under
. Default is 100.0%.
CONTROL DELAY TIME
. . . . . . . . . 00:00 to 99:59 mm:ss
If
Control Delay
or
Both
were selected under the
Delay Option
, this parameter is
enabled upon entering
Deposit
and sets the time for which control will be
suspended and the power maintained at the
Soak 2
power level. Default is 00:00.
ION ASSIST DEPOSIT
. . . . . . . . . . Yes / No
Press the
TOGL
key to move through the choices. Default is
No
.
Yes
activates the
Ion Assist Output
upon entering
Deposit
and deactivates it on leaving
Deposit
.
NAME (XTC/3M only)
. . . . . . . . . . . 15 alphanumeric characters
Accepts up to 15 alphanumeric characters to uniquely identify a film. The film name
can be entered by moving to the
Film Name
and using the keypad in a manner
similar to a cell phone. For instance, to enter the letter S, press the
PQRS
key four
times. To enter the letter A, press the
ABC
key once.
shows a film with
the name:
FILM NAME
.
Names that have been entered via Communications may contain characters
beyond the 15 alphanumeric ones that can be entered from the front panel.