1 - 10
XTC/3 Operating Manual
# of TTL Compatible Outputs . . . . . . 8 standard, Internally pulled up to 5 V (dc).
May be pulled up externally to
24 V (dc) through 2.4 k resistor.
minimum high level 0.5 mA load @3.75 V
maximum low level 10 mA load @1.1 V
Inputs (TTL Compatible). . . . . . . . . . 9 standard
Input Levels
Maximum High . . . . . . . . . . . . . . 24 V
Minimum High. . . . . . . . . . . . . . . 2.5 V
Maximum Low . . . . . . . . . . . . . . 1.1 V
Scan/Update Rate . . . . . . . . . . . . . . 4 Hz
1.4.7 Remote Communications
RS232C Serial Port . . . . . . . . . . . . . Standard; see
Baud Rates. . . . . . . . . . . . . . . . . 9,600; 19,200; 38,400; 57,600; 115,200
Ethernet TCP/IP Port . . . . . . . . . . . . Optional. Static address,
DHCP not supported.
1.4.8 Accessories
Handheld Power Controller Option. . (connects to jank on front panel)
Functions . . . . . . . . . . . . . . . . . . Increase or decrease power, Stop,
CrystalSwitch
Connector Kit . . . . . . . . . . . . . . . . . . Connectors for inputs and relays
Operating Manual. . . . . . . . . . . . . . . CD, PN 074-5000-G1
1.4.9 Power
Power . . . . . . . . . . . . . . . . . . . . . . . . 100 - 230 V (ac), 50/60 Hz
Maximum Apparent Power . . . . . . . . 60 VA max.
Fuse . . . . . . . . . . . . . . . . . . . . . . . . . 1.25 A, 250 V, Type (T)
Temporary Overvoltages . . . . . . . . . Short Term: 1430 V, <5 s
Long Term: 480 V, >5 s