![Inficon Sensistor ISH2000 HySpeed Operating Manual Download Page 50](http://html.mh-extra.com/html/inficon/sensistor-ish2000-hyspeed/sensistor-ish2000-hyspeed_operating-manual_2055023050.webp)
Hydrogen leak detector Sensistor ISH2000
-
Operating Manual
50
OM-Sensistor ISH2000-EN-201304
EN
10 Sensistor ISH2000 specifications
Table 10-1. Power supply specifications.
Table 10-2. Input and output connections.
Power
Sensistor
ISH2000
Sensistor
ISH2000C
Sensistor
ISH2000P
AC mains voltage
100-240 V 50/
60Hz.
100-240 V
50/60 Hz
-
AC mains current
Typically 1 A (2 A
pulse at power
on).
Typically 300 mA
-
Fuse
2 A slow/ 250
VAC.
-
Nominal battery
voltage
-
16.1 VDC
-
Operating time
-
h without screen
saver, at 20 C.
-
Charging time
-
6.5 h
-
Power supply
voltage
-
-
24 VDC
Power supply
current
-
-
3 A max.
Input/output
Sensistor
ISH2000
Sensistor
ISH2000C
Sensistor
ISH2000P
Power input
connector
AC input connector,
IES 320.
Charger input
connector, 2.1 x 5.5
mm std. Positive
centre.
4 pin Phoenix MC 1.5/
5.81 Series
Detachable screw
terminal.
Probe control/Status
port
25-pin D-sub female.
-
25-pin D-sub female.
Minimum pulse length
40 ms
-
40 ms
Input impedance
50k ohm
-
50k ohm
Input maximum range
-34 to +38 VDC
-
-34 to +38 VDC
Input high
> 12.0 VDC
-
> 12.0 VDC
Input low
< 8.0 VDC
-
< 8.0 VDC
Output current
max 0.5 A/output,
max 2.5 A total
-
max 0.5 A/output,
max 2.5 A total
Inductive loads
External clamp diodes
recommended
-
External clamp diodes
recommended
Low state voltage
Max 1.5 VDC
-
Max 1.5 VDC
Low state leakage
current
Max 100 µA
-
Max 100 µA
Short circuit
protection
Thermal and
electronic
-
Thermal and
electronic
Summary of Contents for Sensistor ISH2000 HySpeed