Hydrogen leak detector Sensistor ISH2000
-
Operating Manual
46
OM-Sensistor ISH2000-EN-201304
EN
Installation process
Table 8-13. Install the detector as follows
Fig 8-8. Fixing the mounting brackets.
See “Sensistor ISH2000 specifications” on page 50 for electrical connections
8.18 Default parameters
Table 8-14. Range and default settings of all Sensistor ISH2000 parameters.
Step
Action
1.
Cut hole in panel according to figure above, and
remove burrs.
2.
Check that rubber o-ring is in correct position in
the grove around the edge of the detector.
3.
Put the detector in place in the panel hole.
4.
Hold the panel in place while fixing the mounting
brackets to the detector (See Figure 8-7).
5.
Align the detector horizontally and lock it by
tightening the 4 set screws.
6.
Lock the set screws with the locking nuts.
7.
Connect ground terminal to protective ground of
cabinet.
Bracket to panel screw
Bracket to unit
Bracket
Parameter
Range
Default
APC Time A
0.0 - 6000.0 s
10.0 s
APC Time B
0.0 - 6000.0 s
0 s
APC Time C
0.0 - 6000.0 s
0 s
APC Time D
0.0 - 6000.0 s
0 s
Audio Base Frequency
Several Choices
400 HZ
Auto Range
ON/OFF
ON
Audio Ready Pulse
ON/OFF
ON
Audio Threshold
(Locating)
0 - 100%
0%
Audio Threshold
(Measuring Mode)
0 - 100%
0%