44
tina46e1-b (2017-12) BPG402.om
Insert the pin through the opening shown in the following illustration and
push the button inside for 1 s.
BPG402-S
BPG402-SL
BPG402-SD
BPG402-SE
BPG402-SP
max. ø1.3
Gauges with display will show the
reading "1000 mbar".
Zero point readjustments are automatically carried out during operation of the
gauge, no manual adjustment is needed.
In the event of a fault or a complete failure of the output signal, the gauge can eas-
ily be checked.
•
Voltmeter / ohmmeter
•
Allen wrench, AF 2.5
•
Spare sensor (if the sensor is faulty)
The output signal is available at the sensor cable connector (Pin 2 and Pin 12).
In case of an error, it may be helpful to just turn off the mains supply and
turn it on again after 5 s.
6.2.2 Zero Point Adjustment
6.3 What to Do
in Case of Problems
Required tools / material
Troubleshooting (Gauge)