tina46e1-b (2017-12) BPG402.om
9
BPG402-S, -SL
BPG402-SD, -SE, -SP
Adjustment range
1 ("SETPOINT")
2 ("SETPOINT A, B")
1×10
-9
mbar … 100 mbar
Setpoints adjustable via potentiometers,
one floating, normally open relay contact
per setpoint (
→
19, 20, 39).
(Adjusting the setpoints via field bus
→
corresponding bus section)
Hysteresis
10% of the
threshold value
Relay contact rating
≤
30 V (dc),
≤
0.5 A (dc)
Data rate
Data format
Connections (sensor cable connector)
TxD (Transmit Data)
RxD (Receive Data)
GND
9600 Baud
binary
8 data bits
one stop bit
no parity bit
no handshake
Pin 13
Pin 14
Pin 5
Function and communication protocol of the RS232C interface
→
31
Fieldbus name
DeviceNet
Standard applied
→
[7]
Communication protocol, data format
→
[1], [5]
Interface, physical
CAN bus
Data rate
(adjustable via "RATE" switch)
125 kBaud
250 kBaud
500 kBaud (default)
"P" (125 kBaud, 250 kBaud, 500 kBaud
programmable via DeviceNet)
(
→
[1])
Node address (MAC ID)
(Adjustable via "ADDRESS", "MSD",
"LSD" switches)
0 … 63
dec
(default = 63
dec
)
"P" (0 … 63 programmable via
DeviceNet,
→
[1])
DeviceNet connector
Micro-Style, 5-pin, male
Cable
shielded, special DeviceNet cable,
5 conductors (
→
21 and
[5])
Cable length, system wiring
according to DeviceNet specifications
(
→
[7], [5])
Fieldbus name
EtherCAT
Standard applied, data format,
communication protocol
→
[11], [12]
Data rate
100 Mbps
Node address
explicit device identification
Physical layer
100Base-Tx (IEEE 802.3)
EtherNET connector
2×RJ45, 8-pin, socket
<IN>: EtherCAT input
<OUT>: EtherCAT output
Cable
shielded, 8-pin special Ethernet Patch
cable (quality CAT5e or higher)
Cable length
≤
100 m
Switching functions
RS232C interface
DeviceNet interface
(BPG402-SD)
EtherCAT interface
(BPG402-SE)