Intelligent current clamp meter ISMZ I
– User Manual
10
Display function
Use the arrow buttons (6) and (10) to select DISPLAY OFF.
Use the arrow buttons (7) and (9) to select the time for switching
the display off automatically. The following values are possible:
Always on
1 minute
3 minutes
5 minutes
7 minutes
10 minutes
20 minutes
25 minutes
30 minutes
While recording or observing a current, the clamp will switch into
standby mode after the set time.
Sampling frequency
Use the arrow buttons (6) and (10) to select SAMPLING
Use the arrow buttons (7) and (9) to change the sampling
frequency. The following values are possible:
10 kHz
20 kHz
40 kHz
Changing the sampling frequency also changes the measurable
frequency range. The sampling frequency should always feature at
least the double value of the frequency to be measured. The higher
the sampling frequency, the shorter the battery life.