Procedure
1. Check coolant.
2. Check fluid lines for damage such as buckling and abrasions, and for liquid
leaks. If necessary, replace fluid lines. Record the age or date of replacement
of fluid lines in the service, testing and maintenance schedules or reports.
3.
Cooling water distribution unit with screen pipe (GF Piping Systems)
a
Shut-off valve 1
b
Shut-off valve 2
c
Screen pipe (in plastic or metal, depending on design/variant)
d
Filter cover
e
Valves
f
Flow monitor
Check the filter in the cooling water distribution unit. Close the two shut-off
valves (a and b). Open the filter cover (d) and remove the screen pipe (c).
Clean the screen pipe or replace if necessary and refit. Close the filter cover
again and reopen the shut-off valves (a and b).
4. Check water temperature and flow rate of cooling water (sensor system).
Maintenance Summary - Care activities
33
Maintenance Instructions
MS32-6.2
DIM012EN - 03.06.2020