8.9 

=  Automatic Power Switch Off 

 

The  ADG  balance  includes  an  internal  rechargeable  battery.  The  typical 

operation  time  using  only  the  battery  is  20  hours.  The  balance  includes  the 

TIME function to turn the power off after 5 minutes if the balance is not being 

used.  This  function  can  be  disabled  if  the  balance  is  powered  from  the  main 

power  supply  or  if  the  disruption  of  the  power  might  affect  the  weighing 

procedure. 

 

  To set the parameter press the [Func] key to show “ ”. 

 

  Press [Print] to see the parameter.  

 

  Press [Print] to select either  = 0 or   = 1. 

 = 0  Automatic switch off is enabled. 

 = 1  Automatic switch off is disabled. 

 

  Press the [Func] key to return to weighing. 

 

 

8.10 

 = Hold Maximum Value 

 

 

When  a  weight  is  placed  on  the  pan  the  display  will  hold  the  highest  reading 

until  the  operator  presses  the  [Tare]  key  to  reset  the  display  to  zero.  This 

function  can  be  used  at  the  same  time  as  any  of  the  weighing  functions  i.e. 

Parts Counting, Percent, Newtons etc. When the function is enabled the display 

will show a dash above the left digit. 

 

  To set the parameter press the [Func] key to show “

”. 

 

  Press [Print] to see the current parameter.  

 

  Press [Print] to select 

= 0 or 

 = 1. 

 

 

 = 0 will disable the hold function and  

 = 1 will enable the same. 

 

  Press the [Func] key to return to weighing. 

 

To  exit  this  function  press  the  [Func]  key  until 

  or  any  other  weighing 

function is displayed and then press [Tare]

Summary of Contents for BA165

Page 1: ...User Guide User Guide BA165 Electronic Balance 6000g x 0 1g Impact Test Equipment Ltd www impact test co uk www impact test com...

Page 2: ...t Ltd Building 21 Stevenston Ind Est Stevenston Ayrshire KA20 3LR T 01294 602626 F 01294 461168 E sales impact test co uk Test Equipment Web Site www impact test co uk Test Sieves Accessories Web Site...

Page 3: ...8 2 Digital Filtering Selection 8 8 3 Baud Rate Selection 8 8 4 Parts Counting Function 9 8 5 Calibration Procedure 11 8 6 Results Printed Automatically or Manually 11 8 7 Print when stable or instant...

Page 4: ...m weight The standard RS 232 interface allows the weight to be sent to a computer or printer and allows control of the balance using commands from a PC 2 LOCATING THE BALANCE Place the balance on a st...

Page 5: ...ication scale interval e 1g 1 g 1g Operating temperature 5 C 35 C Internal Battery 12VDC approximately 20 hours operation Power supply 14VDC 100mA minimum Supplied from external power supply 230VAC or...

Page 6: ...supply connector in the plug on the rear of balance Plug the power supply into the mains supply Press the On Off key to activate the balance Wait until the balance auto test is finished and a stable z...

Page 7: ...lance includes a 12 Volt lead acid internal battery as standard This battery will allow approximately 20 hours of operation from a full charge The balance can be charged overnight using the external p...

Page 8: ...ight of the items which is the net weight This may be repeated as many times as necessary to add more items to the container Each time only the weight of the new addition will be shown Make sure that...

Page 9: ...ously or only when stable Enable the auto zero function Enable the automatic power off function Set the balance to hold the maximum value function Display Newtons function A measure of the power suppl...

Page 10: ...h as filling Slowest display rate for applications such as animal weighing or unstable environment 8 2 Digital Filtering Selection Select the required level of digital filtering Press the Func key unt...

Page 11: ...The number of items is the sample size To enter parts counting mode press Func until is displayed Press Print to enter parts counting mode The sample size is displayed To select a different sample siz...

Page 12: ...r weighing function is displayed and then press Tare The diplay will show StEPS The check weighing function will show a symbol on the left side of the display to indicate if the current weight display...

Page 13: ...played initially When is displayed remove the calibration weight The balance will return to normal weighing Verify calibration by placing the weight back on the balance Note The only permissible calib...

Page 14: ...ion The balance has an autozero function to automatically rezero the balance This function will reset the zero of the balance if the zero should drift from the initial zero condition The autozero func...

Page 15: ...ff is enabled 1 Automatic switch off is disabled Press the Func key to return to weighing 8 10 Hold Maximum Value When a weight is placed on the pan the display will hold the highest reading until the...

Page 16: ...he balance can display the voltage from the internal battery To set the parameter press the Func key to show Press Print to see the voltage A fully charged battery will show above 12 Volts If it shows...

Page 17: ...e T not t The balance will send the message ES if it does not understand a command that is sent to it T cr lf The uppercase T will tare the balance This is the same as pressing the Tare key SI cr lf T...

Page 18: ...te correctly Make sure the pan is correctly installed before turning on the balance If the message is still shown after applying power the weighing mechanism might be damaged Displayed when the balanc...

Page 19: ...to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment The equipment generates uses and can radiate radio frequency energy and if not...

Page 20: ...tronic Scales for Medical use Retail Scales for price computing Adam Equipment Co Ltd Bond Avenue Milton Keynes MK1 1SW UK Phone 44 0 1908 274545 Fax 44 0 1908 641339 e mail sales adamequipment co uk...

Reviews: