Model SR565

i

-H Bariatric Stand-On Scale 

Operating and Service Manual 
Part No.  MAN565

i

-H__MICROBAN_180104 

 

Page 1 of 20 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email

[email protected]

     

  

Copyright 2018

 

S

Instruments, Inc. 

 

 

by 

S

 Instruments, Inc. 

 

 

 

 

 

 

 
 
 
 
 
 
 
 

 

Operating and Service 

Manual 

M

odel 

SR

565

i-

H

 

Bariatric Stand-On Scale 

 

 

Содержание SR Scales SR565i-H

Страница 1: ..._MICROBAN_180104 Page 1 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2018 SInstruments Inc S by S Instruments Inc Operating and Service Manual Model SR565i H Bariatric Stand On Scale ...

Страница 2: ... SYSTEM DESCRIPTION AND INTENDED USE 6 MAINTENANCE AND CLEANING 7 STORAGE AND TRANSPORTATION 7 SPECIFICATIONS 8 BUTTON FUNCTIONS 9 BASIC SYSTEM OPERATION 10 BATTERY REPLACEMENT 12 THEORY OF OPERATION 13 CALIBRATION 14 CONFIGURING SEND BUTTON 16 TROUBLESHOOTING 17 WARRANTY 18 TABLE OF FIGURES Figure 1 Scale Assembly Diagram 4 Figure 2 Display Cable Connector and Battery Compartment 4 Figure 3 Print...

Страница 3: ...ents Inc PACKING CHECKLIST DESCRIPTION QUANTITY Box 1 BASE ASSEMBLY 1 ea MAST ASSEMBLY with DISPLAY and HEIGHT BAR 1 ea DISPLAY UNIT WITH CABLE 1 ea PACKAGE OF SIX 6 D CELL BATTERIES 1 ea 3 8 16 FLAT HEAD SCREWS 4 ea CALIBRATION CERTIFICATE 1 ea WARRANTY CARD 1 ea MANUAL 1 ea Box 2 SR8220 Optional Printer Kit PRINTER 1 ea PRINTER BATTERY CHARGER 1 ea PRINTER CABLE 1 ea PRINTER MAST BRACKET HARDWAR...

Страница 4: ...r 3 seats correctly in the Base Assembly STEP 3 Figure 1 Insert the four 4 3 8 16 screws 4 into the holes in Mast Assembly and Base Assembly as indicated Secure the Mast Assembly by tightening the screws into Base Assembly Figure 1 Scale Assembly Diagram STEP 4 Gently rest the scale on its side Verify that the serial number on the bottom of the Display Unit 5 matches that on the Bottom Cover locat...

Страница 5: ...sition INSTALLING OPTIONAL PRINTER STEP 1 Mount the Printer Mast Bracket 8 to the rear of the Mast Assembly 1 using the two 2 mounting screws provided NOTE The three holes on the Printer Mast Bracket should be towards the top of the Mast figure 3 STEP 2 Apply a small amount of Loctite to the three 3 small screws provided and attach the Printer Mounting Bracket to the Printer Mast Bracket figure 4 ...

Страница 6: ...tes the need for an external battery charger or the danger of an AC power supply cord on a portable scale The patient s weight is displayed on a 16 character dot matrix LCD The weight data may be viewed in either pounds or kilograms with a displayed resolution of 0 1 for each INTENDED USE The SR565i H Bariatric Stand On Scale is specifically designed for use as a portable patient weighing system f...

Страница 7: ...leaning the display window as it is made of clear polyester and can be scratched by abrasive cleaners Mild soap and water is recommended for general cleaning and disinfecting STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods longer than three 3 months remove the batteries To maintain proper operation of this instrument storage and transport conditions should not vary outside...

Страница 8: ...isplayed resolution for calibrated range AUTO ZERO One button operation AUTO POWER DOWN Approximately 30 seconds adjustable to 300 seconds RECALL Recalls last stored weight height and BMI AVERAGING Automatic digital filter POWER SUPPLY Six 6 D cell batteries CALIBRATION Calibration is traceable to NIST standards OPERATING CONDITIONS Normal operating conditions for this product Ambient Temperature ...

Страница 9: ...ator flashes then remains solid when stable Auto stores stable weight in memory SEND PRINTER EHR Press to send stored values to EHR or to printer Output values include time date weight height and BMI Also indicated on display when paper is out PAPER OUT and when door is open PAPER DOOR OPEN RECALL Press to recall the last stable weight Display will scroll thru WT HT and BMI stored values in approx...

Страница 10: ...iguring scale output Use the UP or DOWN arrow buttons to select between PRINT and EHR Press ENTER to save changes Setting printout orientation Use the UP or DOWN arrow buttons to select between PRINT INVERTED and PRINT NORMAL Press ENTER to save changes UP Press UP to adjust height up from the default to increase the scale s on time or to select a digit when setting time and date DOWN Press DOWN t...

Страница 11: ...roing the system This will ensure that the patient s NET weight will be displayed It is recommended that the system be zeroed prior to each new patient CALCULATING BMI Press to calculate BMI When the BMI button is pressed the default starting value is set either to 65 in or to 165 cm example HT 65 in Use the UP or DOWN arrows to adjust the default height to the value of the patient s height NOTE T...

Страница 12: ...Tip the scale on its side and unscrew the four 4 screws in the Battery Compartment Cover located on the base of the unit between the wheels STEP 3 Remove and replace ALL six 6 D cell batteries Refer to Battery Compartment Cover label for placement STEP 4 Press the WEIGH button to confirm display is working STEP 5 Tightly close the cover STEP 6 Zero the system Figure 6 Battery Compartment WARNING D...

Страница 13: ...tiometers serving to normalize the sensitivity voltage out per unit of weight applied of each bridge The offset potentiometer produces a small current which nulls the output of the amplifier for an unloaded system The output of the operational amplifier is digitized by the analog to digital converter The sigma delta converter sums a rapid sequence of 0 s 0 volts and 1 s reference voltage to achiev...

Страница 14: ...scale to stabilize note scale reading Remove weight STEP 4 Place second calibrated weight on scale Wait for scale to stabilize note scale reading Remove weight STEP 5 The scale readings for both weights should be within the Calibration Tolerance plus or minus one digit of displayed resolution Figure 7 Figure 7 Calibration Tolerance Table Figure 8 Calibration Switch Diagram CALIBRATION TOLERANCE TA...

Страница 15: ...ht Use the UP arrow button to select the digit and the DOWN arrow button to change digit Press ENTER to save changes STEP 4 Press the MENU button until HALF 500 00 Lb is displayed Set the HALF value of the second selected calibrated weight Use the UP arrow button to select the digit and the DOWN arrow button to change the digit Press ENTER to save changes STEP 5 Press MENU button until 3 Pt Calibr...

Страница 16: ... Records EMR Electronic Health Records EHR technology is supported by the i series scales from SR Instruments The EMR EHR software is not included with scale There are many different EMR EHR software available and the connectivity of the scale to the software requires the services of a professional A null modem cable is required to communicate with the scale Scale output when send button is config...

Страница 17: ...e battery Wait five seconds then re install the battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label is correctly plugged into display board Check to ensure the calibration switch is not in the Calibration Mode Figure 8 The display shows no reading at all Check to ensure batteries are installed co...

Страница 18: ... Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liable for any subsequent or special damages Any misuse improper installation or tampering shall void this warranty DAMAGED SHIPMENTS Title passes to purchaser upon delivery to Tran...

Страница 19: ...ting and Service Manual Part No MAN565i H__MICROBAN_180104 Page 19 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2018 SInstruments Inc NOTES ...

Страница 20: ... No MAN565i H__MICROBAN_180104 Page 20 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2018 SInstruments Inc S by S Instruments Inc Precision Technology in Perfect Balance ...

Отзывы: