![IMKO SONO-DIS Manual Download Page 8](http://html1.mh-extra.com/html/imko/sono-dis/sono-dis_manual_2047641008.webp)
8
3 Initial Commissioning
3.1 Safety Instructions
Attention:
Before initial commissioning, please read the General Notices in the first chapter of this operating instructions.
Any not intended use may lead to damage of the device.
3.2 Checking the scope of delivery
•
SONO-DIS Handheld device
•
Plug-in Power Supply Unit (12V/2A)
•
Charging Adapter
•
Protective Hood
•
Manual
3.3 Charging the Accumulator
The integrated accumulator should be charged before putting the device into operation the first time. For this
purpose, plug in the provided charging adapter into the 7-pole socket of the SONO-DIS. Subsequently, connect
the plug-in power supply unit to the charging adapter. In the event that the device is already switched on, or if the
accumulator is completely discharged, the charging process will commence immediately.
If not, switch on the SONO-DIS by pressing the button „Measurement“
for approximately 1 second. An active
charging process is indicated in the display by an animated accumulator symbol. The integrated charging electronics
charges the accumulator until it is completely charged. In case of complete discharge, this will approximately take
2 hours. As soon as the charging process is completed, all 4 „accumulator bars“ will be permanently presented in
the display and the trickle charging will commence.
Attention:
Only charge the accumulator at room temperature (10°C to 30°C). At too low temperatures, it may happen
that the accumulator is over-charged. Too high ambient temperatures may lead to damage of the SONO-DIS
due to the additional heat-up during the charging process.
3.4 Connecting a Sensor
The SONO-DIS handheld device can be operated with the following IMKO-moisture probes:
Connect the moisture probe to the SONO-DIS by plugging in the 7-pole plug into the socket at the SONO-DIS and
fastening the coupling nut.