42
Menu item: “CHANGE”:
In this menu an own material calibration can be set, or an existing one can be changed according the individual
requirement. Two options are available:
1 point calibration:
- Shifts an existing calibration curve to a choosen
moisture point.
- The gradient isn’t changed.
- Only one measurement is necessary.
2 point calibration
- Creates a linear calibration between two measured
moisture points
- Two material samples with different moisture values
are necessary
1.7.1 1 Point Calibration
In this material calibration option an offset of the appointed calibration is done. As no change of the calibration
curve gradient is proceeded, it is necessary to choose a calibration curve that fits to the material.
NOTE:
To perform a 1-point material calibration you need a sample of the material to be measured. The moisture of
this material has to be exactly determined with another method before starting the calibration.
Procedure
First you have to set the calibration curve to be changed
(01 – 15) by pressing the buttons „Up“
and „Down“
.
Apply your setting by actuating the button „Measure
ment"
.
Next set the moisture of the material by pressing the
buttons „Up“
and „Down“
. Apply your setting
by actuating the button „Measurement“
.