16
ADC-3981 Installation and Operation Manual
Chapter 2: Installation and Removal
2. If the affected module has not been discovered by your CCS
software application, enter the Build mode, and then drag or copy
and paste the module’s device icon from the catalog folder into the
Network
or
Discovery
folder.
3. Right-click the device icon and then select
Properties
.
4. On the
NRO
or
Device
tab of the
Navigation Properties
box, enter
the IP address of the frame that holds the module. (See
Figure 2-3
.)
Figure 2-3.
Navigation Properties Box
5. In the last field, enter the slot number of the module, and then close
the window.
6. Double-click the device icon.
The
Configuration...
box opens. On the
Software Upgrade
tab,
the
/slotx/boot
(where x is the slot number) directory appears in the
Select the device directory to transfer to:
field.
7. Click
Add,
and in the
Add Upgrade Files
box, browse and select
the
boot
folder in the module’s upgrade.
8. Click
OK
.
9. Select the file and then click
OK
.
10. Click
Perform Transfer
and then click
Yes
.
This may take several minutes.
11. Wait for the message
File transfer to device succeeded
in the
status bar.
Caution
Do not make changes in the
third field (located above and to
the right of the
Set Default
button.)
Do not make changes in
this field
Enter frame IP number here
Summary of Contents for neo adc-3981
Page 3: ...Edition B December 2005 ADC 3981 Analog to Digital Converter Installation and Operation Manual...
Page 4: ......
Page 8: ...vi ADC 3981 Installation and Operation Manual Contents...
Page 32: ...18 ADC 3981 Installation and Operation Manual Chapter 2 Installation and Removal...
Page 56: ...42 ADC 3981 Installation and Operation Manual Appendix A Tree View Navigation...
Page 63: ......