8
TORQUE MEASUREMENT
3. While the torque value appears on
the display, press the
–
or
+/C
but-
tons to change the value. Hold
(rather than press) the
–
or
+/C
buttons to change values rapidly.
If no button is pressed for three sec-
onds the display zeros. Press
M
,
–
,
or
+/C
buttons to return.
1. Press and release (do not hold)
P
to
turn on the power.
2. After ‘0’ appears, press and release
the
M
button. The first Go/No Go
preset indicator
1
appears with
torque value and unit.
See pages 5-6 to change units.
PRESS AND RELEASE M TO
DISPLAY THE FIRST PRESET
4. Setting GO/NO GO Presets
Digital torque screwdrivers save five Go/No Go presets. The presets
alert the user with intermittent alerts at 90% of the preset torque
value and continuous alerts at 100% and over. Presets cannot be set
at less than 10% of capacity.
3. Real Time Mode
Set the mode to real time
1
t
(see page 5).
1. Press and release (do not hold)
P
to turn on the power.
2. The display shows ––––, the LED flashes and a beep sounds
(If an error code appears see page 14).
3. Do not apply torque until ‘0’ appears.
4. Apply torque gradually.
5. Torque values are displayed but peak torque values are not
held on the display.
6. Press and hold the
P
button for more than two seconds to
turn OFF the power.
10.00
1
lbf-in
M
12.00
1
lbf-in
M
CHANGE VALUE WITH – OR +/C KEYS