C O M M A N D R E F E R E N C E
Command Reference
7/01
LDC-3700 Series
103
C H A P T E R
4
The TEC:ITE? query returns the value of the measured TEC output current.
Notes
The TEC load current is constantly measured and updated, regardless of the TEC mode of
operation.
This measurement is updated approximately once every 600 msec.
If an external booster is used, the ITE measurement will remain zero, as the internal output section
is disabled in that case.
In local operation, the ITE measured value is determined by pressing the ITE switch in the TEC
DISPLAY area of the front panel, and visually reading the value on the TEC display.
Examples
"TEC:ITE?" −response: 2.43, means the measured TEC output current is 2.430 Amps.
"Tec:Ite?" −response: −3.27, means the measured TEC output current is −3.270 Amps.
‘
The TEC:LIMit:ITE command sets the TEC TE current limit value.
Parameters
An <nrf value> representing the limit value of the TE current, in Amps.
Notes
This value also limits the TEC booster output signal voltage to a value which is proportional to the
TEC limit current (approximately 1 V/A).
Examples
"TEC:LIM:ITE 3.5" −action: the TEC current limit is set to 3.500 amps.
"Tec:Limit:Ite 4.0" −action: the TEC current limit is set to 4.000 amps.
The TEC:LIMit:ITE? query returns the value of the TEC’s TE current limit.
Notes
The ITE limit value is in effect for all modes of TEC operation.
Examples
"TEC:LIM:ITE?" −response: 3.5, means the TEC output current limit is 3.500 amps.
"Tec:Limit:Ite? −response: 4.0, means the TEC output current limit is 4.000 amps.
The TEC:LIMit:THI command sets the TEC high temperature limit value.
Parameters
An <nrf value> which represents the upper bound of the TEC load temperature, in
°
C.
Notes
The THI limit value must be in the range 0 − 199.9
°
C. The THI limit value also becomes the
maximum temperature set point.
The default setting of the TEC outoff enable register forces the TEC output to be shut off if the high
temperature limit is reached. (See the TEC:ENABle:OUTOFF command.)
Examples
"TEC:LIM:THI 100" −action: sets the TEC load temperature limit to 100.0
°
C.
"Tec:Lim:thi 30.3" −action: sets the TEC load temperature limit to 30.3
°
C.
TEC:ITE?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
TEC:LIMit:ITE
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
TEC:LIMit:ITE?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
TEC:LIMit:THI
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com