C O M M A N D R E F E R E N C E
Command Reference
98
LDC-3700 Series
C H A P T E R
4
The TEC:ENABle:COND command sets the status condition enable register of the TEC operations.
These conditions are summarized in bit 1 of the status byte.
Parameters
An <nrf value> whose sum represents the enabled bits:
Notes
The enabled TEC conditions can be read by using the TEC:ENABle:COND? query.
The TEC condition status can be monitored by the TEC:COND? query. If any of the enabled TEC
conditions are true, bit 1 of the status byte register will be set.
The enable registers normally retain their values at power−up (as they were at power−down) unless
the power−on status clear flag is set true (see *PSC, Chapter 3).
Examples
"TEC:ENAB:COND 513" −action: enables the TEC status condition register so that the Output Out
of Tolerance and TE Current Limit conditions will be reported in the status byte register.
"Tec:Enable:Cond #H0DFB" −action: enables the TEC status condition register so that any and all
of the above conditions will be reported in the status byte register.
The TEC:ENABle:COND? query returns the value of the status condition enable register of the TEC
operations.
Notes
The enabled TEC conditions can be set by using the TEC:ENABle:COND command.
The enabled TEC condition status can be monitored by the TEC:COND? query.
Examples
"TEC:ENAB:COND?" −response: 129, means that the TE Module Open and TE Current Limit
conditions may be reported in the status byte register.
"Radix Hex; TEC:Enable:Cond?" −response: #H0EFB, means that any and all of the above
conditions will be reported in the status byte register.
TEC:ENABle:COND
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
1 − TE Current Limit
256 − N/A
2 − Voltage Limit Error
512 − Output Out of Tolerance
4 − N/A
1024 − Output On
8 − High Temperature Limit
2048 − Ready for Calibration Data
16 − TEC Interlock Enable
4096 − N/A
32 − Booster Enable
8192 − N/A
64 − Sensor Open
16384 − N/A
128 − TE Module Open
32768 − N/A
TEC:ENABle:COND?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
1 − TE Current Limit
256 − N/A
2 − Voltage Limit Error
512 − Output Out of Tolerance
4 − N/A
1024 − Output On
8 − High Temperature Limit
2048 − Ready for Calibration Data
16 − TEC Interlock Enable
4096 − N/A
32 − Booster Enable
8192 − N/A
64 − Sensor Open
16384 − N/A
128 − TE Module Open
32768 − N/A