REVEXmax - User’s Manual
4
Measurements
4.8.2
Measuring of power and current consumption by means of current
clamp
WARNING
It is important to observe these principles for proper measurement:
Both the REVEXmax and the tested appliance must be powered from
identical phase.
Procedure:
•
Press the
[9 Power]
key repeatedly in order to select the external method.
The symbol
is displayed.
•
Connect the optional current clamp to the terminal
8
.
•
Switch on the tested appliance’s mains switch.
•
Enclose with current clamp’s jaws live conductor L of the measured appliance.
•
Press the
[TEST]
key.
•
The last measured values are displayed after finishing the measurement.
•
Switch off the tested appliance’s mains switch.
If the measured value of power is higher than the highest measurable value the
symbol
>>>>
appears.