REVEXmax - User’s Manual
2
Description of the instrument
Features and highlights:
•
test results can be saved under an ID number which is unequivocally assigned
to the respective DUT (barcode)
•
the test structures created at the PC software can be loaded into the REVEX-
max
•
measurement data saved to the test instrument can be imported to PC software
via the USB port and reports can be generated
•
user-defined test sequences created at the PC software
•
user-defined report templates created at the PC software
2.2
Terminals
1. Test Socket for connection of tested appliance’s power supply cord
2.
Voltmeter
[
−
] - external voltage test terminal (WELD variant only)
3.
Voltmeter
[+] - external voltage test terminal (WELD variant only)
4.
PE
– terminal for connecting the exposed conductive part of the tested appli-
ance (the terminal is connected to the PE terminal of Test Socket
1
)
5.
I
TOUCH
– touch leakage current test terminal
6.
R
ISO
/
I
ALT
– insulation resistance / alternative leakage current test terminal
7.
R
PE
– earth bond test terminal
8. EXT input - current clamp / three-phase adapter / other adapters connector
9. Fuse F 16 A / 500 V
10. USB - socket type A for connection USB 2.0 Full Speed HID (Human Interface
Device), e.g. barcode scanner
11. USB – socket type B, PC connection
12. Power supply cord
WARNING
•
Do not connect any external voltage to any terminal or connector.
The only exception is the power supply cord
12
which is intended
for powering the REVEXmax from the 230 V / 50 Hz mains
outlet.
•
Max. load on Test Socket
1
is 16 A (limited in time).
•
Use original accessories only.
•
Do not connect any external voltage between R
PE
terminal and
the PE in Test Socket
1
or PE terminal.
c
2019 ILLKO, s.r.o.
Page 10