![IKA ULTRA -TURRAX Tube Drive Operating Instructions Manual Download Page 15](http://html1.mh-extra.com/html/ika/ultra-turrax-tube-drive/ultra-turrax-tube-drive_operating-instructions-manual_624782015.webp)
15
Technical data
Power supply unit
Input
VAC
A
Hz
100 ... 240
0.8
50 ... 60
Output
VDC
Hz
24
24 W.LPS (limited power source)
50 ... 60
Protection class
2 (double insulated)
Tube Drive
Operating voltage
VDC
mA
24
800
Power consumption, normal operation
W
20
Power consumption, standby operation
W
2
Power output
W
17
Drive
EC flat motor
Speed range:
Mode 1
Mode 2
rpm
300 ... 6000, infinitely
adjustable
up to 4000
up to 6000
Speed setting
rotating/pressing knob
Speed display
scale 0 ... 9
Timer: Mode 1
Mode 2
1 ... 29 min
1 ... 59 sec
Display-timer
digital
Perm. ambient temperature
°C
+5 ... +40
Perm. rel. humidity
%
80
Protection to EN 60529
IP 20
Contamination level
2
Overvoltage category
II
Operation at a terrestrial altitude level
m
max. 2000
Dimensions (W x D x H) without tube
mm
100 x 160 x 40
Weight
kg
0.75
Subject to technical changes!
Summary of Contents for ULTRA -TURRAX Tube Drive
Page 2: ......
Page 23: ...23 40 C IKA IVD IVD IKA BMT 20 IKA BMT 50 D IKA...
Page 25: ...25 B 10 1 E D 300 6000 B 00 Start Stop Power 59 6000 Start Stop C Power 29 4000 Power A 2...
Page 29: ...29 IKA IVD IVD IKA BMT 20 IKA BMT 50 IKA 40 C IKA Tube Drive...
Page 49: ...49 IKA IVD IKA BMT 20 IKA BMT 50 IKA 40 C IKA BG...
Page 51: ...51 IKA IVD IVD IKA BMT 20 IKA BMT 50 IKA 40 C IKA EL...
Page 52: ...52...
Page 53: ...53 Note...
Page 54: ...54 Note...
Page 55: ......