Accessories

R 474 / T 653 Telescopic stand:

The stand pillar must always be lightly oiled, so that raising and low-
ering is not made more difficult due to increased friction.

Apart from this, the telescopic stand is maintenance-free.

Cleaning:

- Use only cleaning agents which have been approved by IKA to 

clean IKA devices: these are water (with tenside) and isoproApanol.

- Wear protective gloves during cleaning the devices.
- Before using another than the recommended method for clean-

ing or decontamination, the user must ascertain with IKA that this 
method does not destroy the device.

Maintenance and Cleaning

Spare parts order:

When ordering spare parts, please give:
-  device type.
-  serial number, see type plate.
-  item and designation of the spare part, see 

www.ika.com

.

Repair:

Please send the device for repair only after it has been cleaned 

and is free from any materials which may constitute a health 

hazard.

For repair, please request the “

Decontamination Certificate

”  from  

IKA, or download printout of it from the IKA website 

www.ika.com

.

If you require servicing, return the device in its original packaging. 
Storage packaging is not sufficient. Please also use suitable transport 
packaging.

The stand pillar is inserted into the baseplate so that the guide groove 
is at the back. The black handwheel bolt is then on the right. The 
stand pillar is then secured from below with a hexagon nut, toothed 
lock washer and a large shaped washer.
All four rubber feet are height-adjustable, so the stand can be set up 
so it does not rock, even on uneven floors.
In addition, a hole is provided in the baseplate of the 

R 474 Tele-

scopic stand

, for screwing on a 16 mm diameter support rod. A 

clamp can be fixed on this rod.

Mounting the unit:

The 

RW  47 digital

 agitator unit or the 

T 65 basic / digital 

ULTRA-TURRAX

®

  dispersion unit are placed on the arm of the 

stand and bolted on from below with the bolts supplied. The help 
of an assistant is needed for screwing on the 

T 65 basic / digital 

ULTRA-TURRAX

®

 dispersion unit.

The cable clips included in the delivery package are required for fixing 
the power cable of the unit.

Safety disconnection device

:

The height-adjustable telescopic stand must be fitted with a safe-
ty disconnection device. We recommend installation of the 

SI 400

  

safety limit switch together with the 

SI 472

 / 

SI 474

 clamp support. 

You will find information on their installation in the operating instruc-
tions of the units.

Commissioning (R 474 / T 653 Telescopic stand):

The stand can be raised and lowered by loosening the handwheel 
bolt. The integral gas shock absorber helps to keep the manual force 
needed for this as light as possible. The manual force must be applied 
close to the stand pillar, so that the stand can be raised and lowered 
with only a slight use of force. If the force for raising or lowering is 
applied at a distance from the stand pillar, a turning moment results, 
leading to increased friction on the pillar guide. As a result, the stand 
can only be moved with great difficulty, or cannot be moved at all. 
Besides this, the danger of the stand toppling over is increased.
When the unit is at the correct height, the handwheel bolt is tight-
ened by hand. Check the tightness of the handwheel bolt regularly 
when using the unit, and retighten it if necessary.

Setting up the stand (R 474 / T 653 Telescopic stand):

Summary of Contents for R 472

Page 1: ...Betriebsanleitung DE 3 Operating instructions EN 6 Mode d emploi FR 9 R 472 Floor stand R 474 Telescopic stand T 653 Telescopic stand...

Page 2: ...We declare under our sole responsibility that this product corresponds to the directives 2006 42 EC and conforms with the following standards or normative documents EN ISO 12100 A copy of the complet...

Page 3: ...eachtung des Sicherheitshinweises zu Tod oder schwerer Verlet zung f hren kann Gef hrliche Situation bei der die Nichtbeachtung des Sicherheitshinweises zu Tod oder schwerer Verletzung f hren kann Gef...

Page 4: ...ind die Laufrollen festzustellen Das T 653 Teleskopstativ dient ausschlie lich der Befestigung von T 65 baisc digital ULTRA TURRAX Dispergierger ten Aufstellen des Statives R 472 Bodenstativ Das R hrw...

Page 5: ...Kraftaufwand m glich ist Wird weit von der Stativs ule entfernt die Kraft zum Anheben oder Absenken angebracht entsteht ein Drehmoment und dadurch erh hte Reibung an der S ulenf h rung Dies f hrt daz...

Page 6: ...w hrleistung erstreckt sich nicht auf Verschlei teile und gilt nicht f r Fehler die auf unsachgem e Handhabung und unzureichende Pflege und Wartung entgegen den Anweisungen in dieser Betriebsanlei tun...

Page 7: ...crews and toothed lock washers supplied in accordance with the T 65 basic digital ULTRA TURRAX operating instructions The unit must be mounted in front vertically above the baseplate a different align...

Page 8: ...king Please unpack the device carefully In the case of any damage a detailed report must be sent immedi ately post rail or forwarder The safety of you cannot be guaranteed If the device is operated wi...

Page 9: ...n on uneven floors In addition a hole is provided in the baseplate of the R 474 Tele scopic stand for screwing on a 16 mm diameter support rod A clamp can be fixed on this rod Mounting the unit The RW...

Page 10: ...ble for freight costs The warranty does not cover worn out parts nor does it ap ply to faults resulting from improper use insufficient care or maintenance not carried out in accordance with the instru...

Page 11: ...re soulev sur la t te de guidage du statif t lescopique l aide d une deuxi me personne et fix conform ment au mode d emploi RW 47 digital avec les quatre vis fournies L appareil doit tre mont debout v...

Page 12: ...RTISSEMENT R 472 Pendant le fonctionnement les rouleaux doivent tre bloqu s Le statif t lescopique T 653 sert surtout la fixation des disperseurs T 65 basic digital ULTRA TURRAX AVIS S curisez les r c...

Page 13: ...cilement Si l effort est exerc loin de la colonne de statif un moment de torsion appara t et le frottement sur le guidage de la colonne s accro t Le statif devient alors difficile voire impossible bou...

Page 14: ...ne Les frais de transport restent alors votre charge Accessoires Garantie Carac ristiques techniques La garantie ne s tend pas aux pi ces d usure et n est pas valable en cas de d fauts dus une utilisa...

Page 15: ......

Page 16: ...80 26253 900 eMail info ika in UNITED KINGDOM IKA England LTD Phone 44 1865 986 162 eMail sales england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Wo...

Reviews: