IKA MICROSTAR 15 control User Manual Download Page 6

 Unpacking 

17

16 Correct 

use

Unpacking

/// Scope of Delivery

Unpack the device carefully. 

Any damage should be notified immediately to 

the shipping agent (post office, railway network or logistics company).

Scope of Delivery

 

› MICROSTAR overhead stirrer

 

› power supply (with exchangeable adapters)

 

› extension arm

 

› hexagonal socket screw

 

› hexagon socket offset screw key

 

› shaft protector R 300 (with 2 fasten screws)

 

› chuck key

 

› USB cable

 

H 67.60

 temperature sensor

 

› user guide

 

› warranty card

MICROSTAR control

Warranty

+

+

+

+

+

+

+

+

Correct use

/// Basics

Use

 

› For mixing / stirring liquids from low to medium viscosities  

with various stirring tools.

 

Stand device 

(chuck points down)

Range of use 

(indoor use only)

 

› Laboratories

 

› Pharmacies

 

› Universities

 

› Schools

The safety of the user cannot be guaranteed:

 

› if the device is operated with accessories that are not supplied or  

recommended by the manufacturer.

 

› if the device is operated improperly or contrary to  

the manufacture’s specifications.

 

› if the device or the printed circuit board are modified by third parties.

Summary of Contents for MICROSTAR 15 control

Page 1: ...MICROSTAR control...

Page 2: ...setup Device setup MICROSTAR control 1 Operator panel and display 2 Chuck 3 Extension arm 4 Temperature sensor socket 5 Power socket 6 USB interface 7 Stirring shaft cap What you should watch out for...

Page 3: ...e safety instructions Keep the operating instructions in a place where it can be accessed by everyone Ensure that only trained staff work with the device Follow the safety instructions guidelines occu...

Page 4: ...cross sleeve and anti rotation element for the agitating vessel Working with the device Danger Do not operate the device in explosive atmospheres with hazardous substances or under water The device is...

Page 5: ...out of the vessel because the speed is too high the device is not running smoothly the device begins to move around because of dynamic forces an error occurs Accessories Protect the device and access...

Page 6: ...USB cable H 67 60 temperature sensor user guide warranty card MICROSTAR control Warranty Correct use Basics Use For mixing stirring liquids from low to medium viscosities with various stirring tools S...

Page 7: ...ions corrected This guarantees a constant speed even if the viscosity of the substance being stirred changes Fluctuations in mains voltage within the permitted tolerance range have no effect on the qu...

Page 8: ...ad range however the actual speed does not match the set speed This condition is maintained as long as neither the motor current nor the device temperature exceeds their permissible limit value In thi...

Page 9: ...7 This symbol means the device is communicating with a Wireless Controller or a PC via Bluetooth The symbol no longer appears if no Bluetooth communication is being performed 8 This symbol means the s...

Page 10: ...n the hexangular bolt as required Securing 25 Assembling the overhead stirrer to the stand Cross sleeve Extension arm Stand Screw bolt Check that the stirrer is held in position rmly prior to each use...

Page 11: ...the machine is ready for operation when the mains plug is plugged in If these conditions are not met safe operation is not guaranteed and the machine could be damaged Observe the ambient conditions te...

Page 12: ...lect the Counter Timer key F 2 Select timer key D and edit timer Turn control knob G Setting the value Pressing the control knob G Con rms the set value and switches to the next value hh mm ss Locking...

Page 13: ...t between on off Pressing the control knob con rms the entry and takes you to the next function 2 Edit Beep Rotating the control knob G you can select between on off Pressing the button con rms the en...

Page 14: ...r machine Slave The stirrer machine sends only at the computer s request Even fault indications cannot be sent spontaneously from the stirrer machine to the computer automation system Commands are tra...

Page 15: ...lure rate For cleaning disconnect the main plug Use only cleaning agents which have been approved by IKA to clean IKA devices Dye isopropyl alcohol Construction material water containing tenside isopr...

Page 16: ...ollows in such cases Disconnect the device from the power supply Carry out corrective measures Restart device Error codes 37 Err 04 Cause motor blockage or overload Effect motor off Solutions disconne...

Page 17: ...stand R 1826 Plate stand R 1827 Plate stand R 182 Cross sleeve RH 3 Strap clamp R 300 Stirring shaft protection with 2 screws FK 1 Flexible coupling H 67 60 Temperature sensor H 67 61 Temperature sens...

Page 18: ...lass A 0 15 0 002 ITI Temperature measurement accuracy 0 5 tolerance PT 1000 DIN EN 60751 class A Timer counter function yes Interface USB Permitted on time 100 Nominal voltage 24 VDC Max current 1 30...