Commissioning
31
30 Commissioning
/// Calibration
Notice! ( )
Stirring tool must be removed from the chuck! In calibration mode the chuck rotates
up to max. speed.
The setting functions can be exited at any time by toucing the
standby key (A).
Switch off the device using the standby key
(A)
and follow the steps below:
1.
Pressing and hold down the control knob
(G)
.
› Touching the standby key
(A)
takes you to the additional functions.
Edit
Bluethooth:
› Rotating the
control knob
(G)
you can select between “on / off”.
› Pressing the
control knob
confirms the entry and takes you to the next function.
2.
Edit Beep:
› Rotating the
control knob
(G)
you can select between “on / off”.
› Pressing the button confirms the entry and takes you to the next function.
3.
Torque calibration:
› Pressing the
control knob
(G)
starts the torque calibration.
Warning! ( )
The chuck rotates up to maximum speed! Do not touch rotating parts!
During torque calibration the device can be stopped by pressing
the control knob (G). The device stops the calibration and returns to the
working screen.
The torque calibration is ended when the max. speed is reached. The working
screen appears and the device is ready for operation.