36 Error 

codes

Err. 02

Cause

 

› hardware driving error

Effect

 

› motor off

Solution

 

› contact the service department

Err. 03

Cause

 

› temperature inside of the device is too high

Effect

 

› motor off

Solution

 

› switch off the device and allow it cool down

Error code | Cause | Effect | Solution

/// Repair

Please send in device for repair only after it has been cleaned and  
is free from any materials which may constitute a health hazard.

For repair, please request the “Decontamination Certificate" form from IKA  
or download a print version from www.ika.com.
If you require servicing, return the device in its original packaging. Storage 
packaging is not sufficient. Please also use suitable transport packaging.

Error codes

/// Troubleshooting correctly

If an error occurs, the fault is shown by an error code on the screen.  
Proceed as follows in such cases:

 

› Disconnect the device from the power supply.

 

› Carry out corrective measures.

 

› Restart device.

 Error 

codes 

37

Err. 04

Cause

 

› motor blockage or overload

Effect

 

› motor off

Solutions

 

› disconnect the device from the power supply

 

› decrease the load on the motor and restart device again

Err. 05

Cause

 

› vibration of the device is higher than the permitted value

Effect

 

› motor off

Solutions

 

› disable the vibration dectection

 › Make adjustments to correct the conditions causing  

the excessive vibrations.

Err. 06

Cause

 

› the device fall from assembly

Effect

 

› motor off

Solutions

 

› disconnect the device from the power supply

 

› secure the device to the stand correctly

Err. 08

Cause

 

› speed sensor fault or overload

Effect

 

› motor off

Solution

 › disconnect the device from the power supply

If the actions described fails to resolve the fault or another error code 
is displayed then take one of the following steps:

 

› Contact the service department.

 

› Send the device for repair, including a short description of the fault.

Summary of Contents for MICROSTAR 15 control

Page 1: ...MICROSTAR control...

Page 2: ...setup Device setup MICROSTAR control 1 Operator panel and display 2 Chuck 3 Extension arm 4 Temperature sensor socket 5 Power socket 6 USB interface 7 Stirring shaft cap What you should watch out for...

Page 3: ...e safety instructions Keep the operating instructions in a place where it can be accessed by everyone Ensure that only trained staff work with the device Follow the safety instructions guidelines occu...

Page 4: ...cross sleeve and anti rotation element for the agitating vessel Working with the device Danger Do not operate the device in explosive atmospheres with hazardous substances or under water The device is...

Page 5: ...out of the vessel because the speed is too high the device is not running smoothly the device begins to move around because of dynamic forces an error occurs Accessories Protect the device and access...

Page 6: ...USB cable H 67 60 temperature sensor user guide warranty card MICROSTAR control Warranty Correct use Basics Use For mixing stirring liquids from low to medium viscosities with various stirring tools S...

Page 7: ...ions corrected This guarantees a constant speed even if the viscosity of the substance being stirred changes Fluctuations in mains voltage within the permitted tolerance range have no effect on the qu...

Page 8: ...ad range however the actual speed does not match the set speed This condition is maintained as long as neither the motor current nor the device temperature exceeds their permissible limit value In thi...

Page 9: ...7 This symbol means the device is communicating with a Wireless Controller or a PC via Bluetooth The symbol no longer appears if no Bluetooth communication is being performed 8 This symbol means the s...

Page 10: ...n the hexangular bolt as required Securing 25 Assembling the overhead stirrer to the stand Cross sleeve Extension arm Stand Screw bolt Check that the stirrer is held in position rmly prior to each use...

Page 11: ...the machine is ready for operation when the mains plug is plugged in If these conditions are not met safe operation is not guaranteed and the machine could be damaged Observe the ambient conditions te...

Page 12: ...lect the Counter Timer key F 2 Select timer key D and edit timer Turn control knob G Setting the value Pressing the control knob G Con rms the set value and switches to the next value hh mm ss Locking...

Page 13: ...t between on off Pressing the control knob con rms the entry and takes you to the next function 2 Edit Beep Rotating the control knob G you can select between on off Pressing the button con rms the en...

Page 14: ...r machine Slave The stirrer machine sends only at the computer s request Even fault indications cannot be sent spontaneously from the stirrer machine to the computer automation system Commands are tra...

Page 15: ...lure rate For cleaning disconnect the main plug Use only cleaning agents which have been approved by IKA to clean IKA devices Dye isopropyl alcohol Construction material water containing tenside isopr...

Page 16: ...ollows in such cases Disconnect the device from the power supply Carry out corrective measures Restart device Error codes 37 Err 04 Cause motor blockage or overload Effect motor off Solutions disconne...

Page 17: ...stand R 1826 Plate stand R 1827 Plate stand R 182 Cross sleeve RH 3 Strap clamp R 300 Stirring shaft protection with 2 screws FK 1 Flexible coupling H 67 60 Temperature sensor H 67 61 Temperature sens...

Page 18: ...lass A 0 15 0 002 ITI Temperature measurement accuracy 0 5 tolerance PT 1000 DIN EN 60751 class A Timer counter function yes Interface USB Permitted on time 100 Nominal voltage 24 VDC Max current 1 30...