29
092008
Check whether the voltage specified on the type plate matches
the mains voltage available. The power socket used must be ear-
thed (protective earth conductor contact). Insert the temperature
measuring sensor (not included in the supply schedule) into the 5-
polar DIN-bushing at the rear of the instrument. Put a magnetic bar
into a vessel that is permeable for magnetic field lines, and fill the
vessel with the medium to be processed. Put the vessel on the
heating plate and immerse the temperature measuring sensor into
the medium (min. 20 mm). See chapter 11.3 for the operation
without external temperature measuring sensor. If these conditi-
ons are met, the device is ready to operate after plugging in the
mains plug. If these procedures are not followed, safe operation
cannot be guaranteed and/or the equipment may be damaged.
Observe the ambient conditions (temperature, humidity, etc.)
listed under Technical Data.
Foil key Mot
Foil key Temp
LED Mot
Foil key Mode
Digital display
Foil key
Visc
LED Temp
Rotary knob Mot
Rotary knob Temp
Commissioning
5
Control elements
6
When the functions are switched off, the digital display always
shows the set speed value.
Mains switch
Display
7
Fig. 1
Real value
Rated value
e.g. phys. unit,
mode of operation,
error
code
Mode of
operation
B or C
Process value; (real value)
function started
Set value
(rated value)
PT 100 or
PT 1000
PC-control
In all three modes of operation only function “Heating” has been
started, the display shifts to rated and real temperature.
In all three modes of operation only function “Stirring” has been
started, the display shifts to rated and real speed.
3386500a_RETcv_092008:RET contr visc_04 06.10.2008 13:43 Uhr Seite 29