8.2 Mixing tool
8.2.1 Removing the mixing tool
•
Loosen feed lines
•
Loosen 4 cap nuts (720) and remove
•
Pull up the inlet flange (704) sidewards
•
Pull up the stator (1006)
•
Loosen screws (424) and replace protection plate (422)
•
Place retaining pin
∅
7 in the bore hole of the shaft (426) by passing it through an
opening in the louver
•
Unscrew the fixing screw (730) with the open-end wrench by turning it counter
clockwise
•
Pull off the rotor (1004) along the shaft
To protect the surface of the tools, please cover sharp edges of the assembly
tool.
8.2.2 Assembly of the mixing tool
•
The assembly is done in the reverse sequence of the assembly as shown in 8.2.1
•
To make disassembly easy coat the hubs of the tools lightly with grease
•
During assembly insure that the O-rings of each tool are placed in the groove cor-
rectly.
Info
The fixing screw (730) should be turned and then tight-
ened (tightening moment of approx. 30 Nm). It is recom-
mended to additionally secure the assembly thread with
a liquid screw retention agent (which can be removed
using standard tools) (e.g. LOCTITE screw retention
agent).