3
Operation Methods
1. Preparing for use
Using the PACK LEADER with batteries
The batteries are mounted before shipment, so the PACK LEADER can be used right away.
When replacing the batteries, use AA alkaline batteries or nickel-metal hydride batteries (ENELOOP).
Using the PACK LEADER with 100 VAC power
Always use the enclosed AC adapter.
•
Always remove the batteries when using the AC adapter or when suspending the use of the
instrument for a long time and putting the unit into storage. The instrument could be damaged
if the battery fluid leaks.
•
Always use the enclosed AC adapter. Failure to do so could damage the instrument.
2. Preparing for measurement
Power ON/OFF
* After turning the power OFF, wait at least two seconds before turning it ON again.
•
The calibration or measured value could be affected if the ambient temperature changes. In
this case, turn the power OFF with the oxygen sensor set, and wait at least 30 minutes for
the temperature to stabilize.
•
Always remove the tip of the needle from the sample before turning the power OFF.
1.
Press
.
The power turns ON, and the software version displays.
(Approx. 4 seconds)
* The displayed details will vary according to the version.
The Measurement Standby screen opens.
* The screen on the left shows the Auto mode and Concentration
% O
2
setting.
2.
Press
again.
The atmosphere is automatically suctioned. After the
screen on the left opens, the power turns OFF.
Sensor warranty term notification function
(Displayed when the self-diagnosis function determines that
there are 30 days or less to the end of the oxygen sensor warranty term.)
Display when 25 days are remaining Display when 0 days are remaining (end of warranty term)
* The warranty is valid for one year from the date of shipment.
(When the sensor is replaced, one year from the date the new sensor
package is opened.)
* The display on the left is a guideline.
The self-diagnosis results are also displayed.
Caution
Note