18
Initialization [INITIALIZE]
Use this function if the measured value is incorrect after calibration. It resets the
O
2
display setting, O
2
calibration value, CO
2
disable setting, upper limit setting,
CO
2
calibration value, measurement mode, and output setting to the factory
default settings. The initialization will reset any arbitrary settings to the factory
default settings.
1.
When the Measurement Standby screen is displayed, press
.
The MENU screen opens.
Immediately after opening,“CAL” is selected (flashes).
2.
Press
twice.
The cursor moves, and “FUNC” is selected (flashes).
3.
Press
.
The FUNC MENU screen opens.
Immediately after opening, “CLOG CHECK” is selected
(flashes).
4.
Press
.
The cursor moves, and “INIT” is selected (flashes).
5.
Press
.
The INITIALIZE screen opens.
6.
Press
.
The INITIALIZE complete screen displays for approx. 1 second, completing
initialization.(The screen returns to FUNC MENU.)
Initialization returns the O
2
span and zero calibration values and the CO
2
calibration value to the factory default
values. The O
2
display setting is reset to “CONC %O
2
”, the CO
2
disable setting to “OFF”, the upper limit setting to “ON”,
the measurement mode to “Auto mode”, and the output setting to “NO OUTPUT”.