3. Then the cuff will be slowly inflated. The blood pressure and pulse will be measured during inflation

Inflation will stop as soon as the blood pressure and pulse rate have been calculated and displayed on

the screen. The result will be automatically stored in the memory, and all results will be uploaded to the

App automatically upon the next successful Bluetooth connection.

SYNC TIME AND RESULTS VIA BLUETOOTH

Connect to iOS Device via Bluetooth

• Press the START/STOP button for 3 seconds, the

Bluetooth

indicator will begin flashing.

• Enable the Bluetooth on your iOS device(Setting->General->Bluetooth->On) .

• Launch the "iHealth MyVitals" app to initiate the connection, the Bluetooth indicator will light up and stop flashing when a

successful connection is established.

• When sync is processing, the sync indicator flash

and

, when sync is finished, the sync indicator lights up

.

Connect to Android Device via Bluetooth

• Press the START/STOP button for 3 seconds, the

Bluetooth

indicator will begin flashing.

• Enable the Bluetooth on your Android Device.

• When using the monitor for the first time, you should pair the monitor to the Android device. Go into your device’s setting to

pair it with your Android device. Check the Bluetooth menu for the model name of your monitor (BP7SXXX) to appear, and then

select it to pair. This may take up to 30 seconds and your monitor’s Bluetooth indicator will light up when a connection has been

established.

• Launch the “iHealth MyVitals” app to sync with your monitor.

• When sync is processing, the sync indicator flash

and

, when sync is finished, the sync indicator lights up

.

• Please repeat these steps when you switch to another Android device with the monitor.

Remember to:

1. Make sure your wrist size is within cuff circumference; see the range in “SPECIFICATIONS”

2. Measure on the same wrist each time.

3. Stay still and calm for one to one and half minutes before taking a blood pressure measurement. Prolonged over-inflation of

the bladder may cause bruises of your wrist.

4. Keep the cuff clean. Cleaning the cuff after every 200 times of usage is recommended. If the cuff becomes dirty, clean it with

a moistened cloth. Do not rinse the monitor or cuff with running water.

5. Press the “START/STOP” button at any time to interrupt a measurement.

Note: Physical activity, eating, drinking, smoking, excitement, stress, and many other factors influence blood pressure results.

SPECIFICATIONS

1. Product name: iHealth View Wireless Blood Pressure Wrist Monitor

2. Model: BP7S

3. Classification: Internally powered; Type BF applied part; IP22, No AP or APG; Continuous operation

4. Machine size: approx. 2.8"×2.9"×0.7" (72mm×74mm×17.6mm)

5. Cuff circumference: 5.3"- 8.7" (13.5cm-22cm)

6. Weight: approx. 4.2oz(120g)(including cuff)

7. Power: DC:5.0V

1.0A, Battery: 1*3.7V

Li-ion 400mAh

Summary of Contents for View

Page 1: ...T UP REQUIREMENTS 2 SET UP PROCEDURES 3 MEASUREMENT PROCEDURES 3 SPECIFICATIONS 4 GENERAL SAFETY AND PRECAUTIONS 5 BATTERY HANDLING AND USAGE 6 TROUBLESHOOTING 6 CARE AND MAINTENANCE 7 WARRANTY INFORM...

Page 2: ...gned to measure the systolic and diastolic blood pressures and pulse rate of an adult individual by using a technique in which an inflatable cuff is wrapped around the wrist The measurement range of t...

Page 3: ...sure measurements 1 Sit comfortably with your feet flat on the floor without crossing your legs Stay still during measurement Do not move your wrist body or the monitor 2 Place your hand palm side up...

Page 4: ...ar and then select it to pair This may take up to 30 seconds and your monitor s Bluetooth indicator will light up when a connection has been established Launch the iHealth MyVitals app to sync with yo...

Page 5: ...ethod within the limits prescribed by the American National Standard Electronic or Automated Sphygmomanometer This device is also clinically validated according to the 2010 Protocol of the European So...

Page 6: ...ery when the battery is less than 25 charged Overcharging the battery may reduce its lifetime When in charging mode the charging status will be displayed on the LED screen See the table below for deta...

Page 7: ...gnetic interference Press the START STOP button and hold for about 10 seconds to reset the device CARE AND MAINTENANCE 1 If this monitor is stored near freezing temperatures allow it to return to room...

Page 8: ...lue The sign graphical symbol white Symbol for ENVIRONMENT PROTECTION Waste electrical products should not be disposed of with household waste Please recycle where facilities exist Check with your loc...

Page 9: ...easures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to which the receive...

Page 10: ...c discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be a...

Page 11: ...all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellul...

Page 12: ...rs m can be determined using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufactu...