PM11
Electronic pressure sensor
8
MEW
MAW
AEP
ASP
[% MEW]
P
-1,25
0
100 103,13
4
3,8
20
20,5
21,5
[mA]
1
2
4
UL
OL
3
5
Fig. 1: Output characteristics of the analogue output to Namur
1:
Analogue signal
2:
Measured value
3:
Measuring range
4:
Scaled measuring range
5:
Signalling fault 21.5 mA
P:
Pressure
MAW:
Initial value of the measuring range with non-scaled measuring range.
MEW:
Final value of the measuring range with non-scaled measuring range
ASP:
Analogue start point with scaled measuring range
AEP:
Analogue end point with scaled measuring range
UL:
Below the display range
OL: