![IFM PM11 Operating Instructions Manual Download Page 12](http://html1.mh-extra.com/html/ifm/pm11/pm11_operating-instructions-manual_4131918012.webp)
PM11
Electronic pressure sensor
12
2
3
4
1
u
Avoid soiling and moisture during the exchange
u
Clean the thread carefully and without residues.
u
Do not damage the adhesive surface for the diaphragm on the sensor.
u
Observe the orientation of the filter cover (
Ò
Installation instructions E30139 / E30467).
When using the closed filter cover, there is no pressure compensation of the measuring cell any
more. Therefore measurement deviations due to:
Fluctuations of the atmospheric pressure.