![background image](http://html.mh-extra.com/html/ifm/ax460/ax460_operating-instructions-manual_620767012.webp)
Operating instructions AX460
12
7.6 Analogue output (DX2043, DX2053)
On terminals 13 and 14 / 15, a 16 bit analogue output is available�
This output can be configured and scaled in the ANALOG MENU� The following
configuration is possible:
●
Voltage output: -10 … +10 V
●
Voltage output: 0 … 20 mA
●
Voltage output: 4 … 20 mA
The analogue output is proportional to the displayed value and refers to the AGND
potential�
AGND and device GND are internally connected�
Parallel operation of voltage and current output is not permitted�
7.7 Control outputs (DX20x2, DX20x3, DX20x5)
On terminals 19, 20, 21, 22 and 23, four control outputs are available�
The switching conditions can be set in the PRESELECTION MENU�
The outputs Ctrl� Out 1 – Ctrl� Out 4 are fast PNP outputs with a switching capacity
of 5 – 30 V / 200 mA per channel�
The switching voltage is determined by the voltage at terminal 19 (Com+)� To
switch inductive loads, external damping measures are recommended� The
switching status is indicated in the display as C1 – C4 (display with unit and status
line)�
Connection of the control outputs