ZSSC4151Evaluation Kit Start-Up Guide
© 2016 Integrated Device Technology, Inc.
9
March 11, 2016
3
Performing a Sample Calibration
This section describes the basic operations to perform a simple, two-point linear calibration at a single tempera-
ture. Higher order calibrations are performed similarly, but with additional points taken over the bridge measurand
and temperature.
3.1. Setting up the Calibration and Acquiring Data
1. Select the “Calibration” tab to display the “Calibration” window (see Figure 3.1).
2. Ensure that the “Calibration Type Settings” and “Acquisition Settings” are set as shown in Figure 3.1. The
“Calibration Table” shows that there are only two points required to complete the calibration in this case.
3. Set the upper and lower “Target” limits to 90% and 10%.
DO NOT
change the “Tgt.Min” and “Tgt.Max”
fields.
Figure 3.1 Using the Calibration Window Tab to Set Upper and Lower Target Limits