Operator Manual ilex
55
L280-0975 REV A
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
Table 4 per 5.2.2.2 of IEC 60601-1-2: 2007.
Guidance and manufacturer’s declaration –electromagnetic immunity
The ILEX55 is intended for use in the electromagnetic environment specified below. The customer or the user of the ILEX55
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Complia
nce level
Electromagnetic environment
–
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
outside ISM bands
a
10 Vrms
150 kHz to 80 MHz
in ISM bands
a
10 V/m
80 MHz to 2.5 GHz
[
V
1
] V
[
V
2
] V
[
E
1
] V/m
Portable and mobile RF communications equipment should be used
no closer to any part of the [
ME EQUIPMENT
or
ME SYSTEM
], including
cables, than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance
┌ 3.5┐ ___
d =
│──│ √
P
└
V
1
┘
┌ 3.5┐ ___
d =
│──│ √
P
└
V
2
┘
┌ 12 ┐ ___
d =
│──│ √
P
80 MHz to 800 MHz
└
E
1
┘
┌ 23 ┐ ___
d =
│──│ √
P
800 MHz to 2.5 GHz
└
E
1
┘
where
P
is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and
d
is the
recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the ILEX55 is used exceeds the applicable RF
compliance level above, the ILEX55 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the ILEX55.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.