• Like Cleansafe® optical cleaning liquid
Do not use following products:
• Alcohol/solvents at higher concentration > 70%
• Strong alkalis lye, strong solvents
• Acid
• Detergents with fluoride
• Detergents with ammonia at higher concentration > 1.6%
• Detergents with abrasives
• Steel wool
• Sponge with abrasives
• Steel blades
• Cloth with steel thread
Please contact Image Diagnostics for Service and Repair at (978) 829-0009 or Toll Free
(877) 304-5434.
10.
Maintenance
All maintenance procedures should be done by an experienced and qualified technician with
demonstrated knowledge and skills (electrical and mechanical) in the service of medical
equipment.
✓
This individual must have access to this manual and the proper tools.
✓
Lubrication of this device is
not
required.
RECOMMENDED PERIODIC PERFORMANCE CHECKS
Daily (Prior to use)
Inspect all external cables, castors and
controls for wear and damage. Replace
damaged cables promptly.
Semi-annually
Check battery operation by disconnecting the
AC power and running the Monitor up and
down.
Annually
Inspect movement of the monitor to ensure
that it is smooth and continuous. Inspect
9.