7.
Batteries
The ilex55 is designed to be used when powered with a hospital grade AC outlet. Battery
backup to elevate the monitor is available for emergency or transportation only. When AC
power is applied, the actuator backup battery will begin charging until the battery reaches full
capacity at which time, the charging will taper off. If the unit loses AC power, the monitor
elevation function will automatically switch to battery back-up mode. An audible tone is
sounded when using an operator control function if the battery charge is low.
Batteries are located inside the control box.
Local procedures may be in place for proper disposal or recycling of batteries.
For more information about batteries, such as battery life, production codes and disposal, see the
Linak
website.
Battery (12V/1.2Ah)
Part Number K000-0040 (Replace in pairs)
Note:
The battery will not charge if the system is not plugged in or the green toggle
switch/thermal breaker is off.
WARNING
Actuator Batteries are
not
automatically maintained to
original maximum charge requiring periodic testing
and replacement if necessary.
CAUTION
Batteries are not to be changed by the operator.
Refer to authorized technician.