IDEXX VetLab
UA
Analyzer
Operator’s Guide
7
Preface
Safety Precautions
The analyzer does not contain any user-serviceable components. DO NOT disassemble.
DO NOT stack other equipment or containers on top of the analyzer.
Keep analyzer away from sources of heat or flames.
DO NOT place or operate the analyzer near x-ray equipment, photocopiers, or other devices that
generate static or magnetic fields.
PROTECT your equipment from damp conditions or wet weather.
Take care not to spill water or other fluids on the unit.
DO NOT use any of the following liquids, abrasives, or aerosol sprays on or near the analyzer, as
they may damage the outer case and may influence results:
Organic solvents
t
Ammonia-based cleaners
t
Ink markers
t
Sprays/disinfectants containing volatile liquids (except isopropyl alcohol)
t
Insecticides
t
Polish
t
Room freshener
t
Use
only
the power adapter supplied (which is marked as supplying 7.5 V DC, 3000 mA). The
cords from other power adapters may fit in the power socket of the UA analyzer, but may have
the incorrect voltage or current for this analyzer. Using a different power adapter than the one
supplied may impact results or render the analyzer unusable.
Disconnect the power cable:
If the cable becomes frayed or otherwise damaged.
t
If anything is spilled onto the equipment.
t
If your equipment is exposed to excessive moisture.
t
If your equipment is dropped or the case has been damaged.
t
If you suspect that your analyzer needs service or repair.
t
Whenever you clean the case.
t
If the equipment is used in a manner other than specified, the protection provided by the
equipment may be impaired.