34
IDEXX VetLab
UA
Analyzer Operator’s Guide
Notification
Cause
Action
Values obtained are different
when compared with those from
visual evaluation
The test strip is not properly
positioned on the test strip tray.
Repeat the test using a new
IDEXX UA strip. Ensure the
test strip is properly aligned in
the middle of the tray before
processing.
For more information, see
“Reading IDEXX UA Strips” on
page 15.
The urine on the test strip has
dried.
Repeat the test using a new
IDEXX UA Strip.
For more information, see
“Reading IDEXX UA Strips” on
page 15.
Strong urine color.
The analyzer uses the
compensation pad on the
IDEXX UA Strip to compensate
for the strong color. Since this
is not possible to do visually,
the analyzer’s reading is more
accurate.
No Printout
The printer is not powered on.
Change the printer setting to
On. For instructions, see “Printer
Settings” on pages 20–21.
No paper has been inserted or
the paper roll is finished.
Load or replace the printer
paper. For more information, see
“Replacing the Printer Paper” on
page 27.
For assistance, contact IDEXX Technical Support (see “IDEXX Technical Support Contact
Information” on pages 37–38).